Calibration
Calibration of Infrared Thermal Camera Using the Black Box Method
Applies To: Infrared Thermal Imaging Devices (e.g., FLIR)
Calibration Method: Black Box (Thermal Reference Source)
Standard Reference: NATA Certification or Traceable Calibration
Permitted Tolerance: ±1°C
1. Purpose
To ensure accurate and reliable thermal readings from infrared cameras used in food safety monitoring by performing a three-point calibration using a thermal black box.

2. Required Equipment
Infrared Thermal Camera (e.g., FLIR)
Certified Thermal Black Box
Preferably NATA-certified (three-point calibration)
If not available, must be traceably calibrated
Tripod or camera mount
Calibration log or digital record
3. Calibration Standard
Perform a three-point calibration at the following temperatures:
50°C – lower operational limit
75°C – for mid-range verification
100°C – upper operational limit
Allowable variance: ±1.0°C at each point
4. Calibration Procedure
Step 1: Preparation
Power on the black box and allow it to stabilize at each calibration temperature.
Mount the infrared camera securely on a tripod or level surface.
Set the camera’s emissivity setting to match the black box surface (typically 0.95 unless otherwise specified).
Step 2: Calibration at 50°C
Point the camera at the 50°C surface of the black box from approximately 0.3 to 0.5 meter.
Allow the reading to stabilize and record the temperature displayed.
If the reading is outside ±1.0°C, adjust the camera per its internal calibration options or document as out of tolerance.
Step 3: Calibration at 75°C
Repeat the same process for the 75°C zone.
Record the reading and confirm it is within the allowed deviation.
Step 4: Calibration at 100°C
Repeat the same for 100°C.
Log the value and verify the result is within the specified tolerance.
5. Documentation
Record the following for each calibration point:
Set point temperature
Measured camera reading
Difference (deviation)
Whether adjustment was required
Date and technician initials
Use your organization’s calibration form or the Didge platform logging module.
6. Notes and Considerations
NATA certification of the black box is preferred. If not feasible, ensure the device has traceable calibration and retain certificates.
Recalibrate the camera:
Every 6 months (recommended)
After repair or accidental drop
If readings are suspected to be inaccurate
Perform calibration in a stable environment without airflow or reflective interference.
The thermal imagery units emissivity value must match that of the blackbox calibration unit
MyFlir app currently has set emissivity values as follows:
Matte e=0.95
Semi-matte e=0.90
Semi-glossy e=80
Glossy e=60
There are no current options for further emissivity settings.
7. Acceptance Criteria
All three calibration points must read within ±1.0°C of the set point.
If any point fails to meet criteria:
Attempt recalibration
If unresolved, flag the camera for service or decommissioning
Last updated
Was this helpful?
