Calibration

Calibration of Infrared Thermal Camera Using the Black Box Method

Applies To: Infrared Thermal Imaging Devices (e.g., FLIR)

Calibration Method: Black Box (Thermal Reference Source)

Standard Reference: NATA Certification or Traceable Calibration

Permitted Tolerance: ±1°C


1. Purpose

To ensure accurate and reliable thermal readings from infrared cameras used in food safety monitoring by performing a three-point calibration using a thermal black box.


2. Required Equipment

  • Infrared Thermal Camera (e.g., FLIR)

  • Certified Thermal Black Box

    • Preferably NATA-certified (three-point calibration)

    • If not available, must be traceably calibrated

  • Tripod or camera mount

  • Calibration log or digital record


3. Calibration Standard

Perform a three-point calibration at the following temperatures:

  • 50°C – lower operational limit

  • 75°C – for mid-range verification

  • 100°C – upper operational limit

Allowable variance: ±1.0°C at each point


4. Calibration Procedure

Step 1: Preparation

  • Power on the black box and allow it to stabilize at each calibration temperature.

  • Mount the infrared camera securely on a tripod or level surface.

  • Set the camera’s emissivity setting to match the black box surface (typically 0.95 unless otherwise specified).

Step 2: Calibration at 50°C

  • Point the camera at the 50°C surface of the black box from approximately 0.3 to 0.5 meter.

  • Allow the reading to stabilize and record the temperature displayed.

  • If the reading is outside ±1.0°C, adjust the camera per its internal calibration options or document as out of tolerance.

Step 3: Calibration at 75°C

  • Repeat the same process for the 75°C zone.

  • Record the reading and confirm it is within the allowed deviation.

Step 4: Calibration at 100°C

  • Repeat the same for 100°C.

  • Log the value and verify the result is within the specified tolerance.


5. Documentation

Record the following for each calibration point:

  • Set point temperature

  • Measured camera reading

  • Difference (deviation)

  • Whether adjustment was required

  • Date and technician initials

Use your organization’s calibration form or the Didge platform logging module.


6. Notes and Considerations

  • NATA certification of the black box is preferred. If not feasible, ensure the device has traceable calibration and retain certificates.

  • Recalibrate the camera:

    • Every 6 months (recommended)

    • After repair or accidental drop

    • If readings are suspected to be inaccurate

  • Perform calibration in a stable environment without airflow or reflective interference.

  • The thermal imagery units emissivity value must match that of the blackbox calibration unit

    • MyFlir app currently has set emissivity values as follows:

      • Matte e=0.95

      • Semi-matte e=0.90

      • Semi-glossy e=80

      • Glossy e=60

    • There are no current options for further emissivity settings.


7. Acceptance Criteria

  • All three calibration points must read within ±1.0°C of the set point.

  • If any point fails to meet criteria:

    • Attempt recalibration

    • If unresolved, flag the camera for service or decommissioning

Last updated

Was this helpful?